WhaleTeq SECG 4.0 Changelog

Version 5.0.14.1 - 2026-01-15

  1. FIX: incorrectly check assistant license

Version 5.0.13.5 - 2025-08-26

  1. FIX: BPM of pacing and waveform are not synchronized
  2. FIX: incorrect waveform type description in 201.12.4.4.108.a
  3. FIX: fail to output signals in YY1079 4.2.8.8
  4. CHG: added 15/ 300/ 350 BPM options in Auto Heart Rate
  5. CHG: added 250 BPM to Superimpose 0.1mV mains frequency noise in 201.12.1.101.15
  6. CHG: update the live update repository URL

Version 5.0.13.4 - 2025-04-16

  1. CHG: removed reference to IEC 60601-2-26 standard

Version 5.0.13.3 - 2025-03-24

  1. NEW: output raw data can set loop or not

Version 5.0.13.1 - 2024-11-13

  1. FIX: incorrect pass criteria of IEC60601-2-27 201.12.1.101.5 in Simplified Chinese version
  2. FIX: fail to download physionet database

Version 5.0.12.3 - 2024-05-12

  1. FIX: crash when outputting signals for a long time
  2. FIX: fail to output signals after the device is re-connected

Version 5.0.11.3 - 2024-02-15

  1. FIX: Assistant YY1079 test item 4.2.8.8 a) 2) - dynamic range test is not activated

Version 5.0.11.2 - 2023-09-03

  1. FIX: incorrect P wave amplitude in CAL05000/ CAL10000/ CAL20000

Version 5.0.10.10 - 2022-10-12

  1. FIX: incorrectly check tool version for live update functionality

Version 5.0.10.9 - 2022-10-06

  1. CHG: update assistant test description

Version 5.0.10.8 - 2022-10-05

  1. CHG: update assistant test description

Version 5.0.10.7 - 2022-09-26

  1. FIX: incorrect amplitude when variable dc is enable

Version 5.0.10.6 - 2022-09-20

  1. CHG: update assistant test description

Version 5.0.10.5 - 2022-07-20

  1. NEW: added support new assistant - GB9706.225/ GB9706.227/ YY9706.247
  2. FIX: (IEC60601-2-25) the initial DC offset is not configured to the device
  3. FIX: incorrect pacing duration when double pacing pulse is used
  4. FIX: (IEC60601-2-27) incorrect test procedure description in 201.12.1.101.14

Version 5.0.10.1 - 2021-08-26

  1. CHG: change the resolution of the pacing overshooting amplitude from 0.1mV to 0.01mV

Version 5.0.9.19 - 2020-10-12

  1. Refine auto pacing UI widget behavior.

Version 5.0.9.18 - 2019-02-13

  1. Non-ISP EEG be recognized by SECG AP

Version 5.0.9.17 - 2018-12-21

  1. FIX: the pacing overshooting is not changed when the pacing overshooting amplitude is updated

Version 5.0.9.16 - 2018-11-28

  1. Enhanced function of Auto-HR /Auto-Pacing
  2. SECG assistant: IEC60601-2-27 test item 201.12.1.101.1 modify. Add steps, let the user input the full scale number before starting the test
  3. SECG assistant: IEC60601-2-27 test item 201.12.1.101.2 modify. Add test Waveform, Triangle or Sine
  4. SECG assistant: IEC60601-2-27, test item 201.12.1.101.16 modify. Add steps, let the user input the full scale number before starting the test
  5. SECG assistant: IEC60601-2-25, 201.12.4.107.1.1.1 modify. Add Method B) C) D)
  6. SECG assistant: IEC60601-2-25 test item 201.12.4.108.3.2 modify. Method B: triangle pulses 1 mV, 50 ms, 120BPM
  7. SECG assistant: IEC60601-2-25 test item 201.12.4.109 modify. Test 1: Pacing add "Off (Vref)" (Pulse Amplitude is zero). Test 2: add "Pulse, 2 mV, width 100 ms"
  8. Calibration mode: Variable DC Offset Limit from 5% modify to 1%

Version 5.0.0.5 - 2016-08-09

  1. Solution for the inaccurate pacing duration while playing double pulse (and single pulse with 0.3/0.6 ms duration).
  2. Enhance SECG Assistant.

Version 5.0.0.4 - 2016-04-19

  1. Assistant I: Change the T-Wave amplitude in -2-27:2011, 201.12.1.101.15 from 0.2 mV to 0mV.
  2. Support to play decimal raw data. (e.g. Sampling Rate = 100.582 Hz.)
  3. Support Hot Plug

Version 5.0.0.3 - 2016-01-26

  1. SECG Assistant: Add new field let user enter specific testing time under the test requirement - IEC 60601-2-27, 201.12.1.101.11 Baseline reset.
  2. SECG Assistant: Solve the issue for "System will crash after inputing 0mV under the test requirement - IEC 60601-2-27, 201.12.1.101.17 Tall T-wave rejection capability".

Version 5.0.0.2 - 2015-11-24

  1. Auto Pacing: Add new functions - "Select All", "Clear All", "Backward", and "Afterward".
  2. Auto Pacing: Solve the issue for drawing wrong waveform after using "Backward" and "Afterward" functions.