WhaleTeq SECG 4.0 Changelog
Version 5.0.14.1 - 2026-01-15
- FIX: incorrectly check assistant license
Version 5.0.13.5 - 2025-08-26
- FIX: BPM of pacing and waveform are not synchronized
- FIX: incorrect waveform type description in 201.12.4.4.108.a
- FIX: fail to output signals in YY1079 4.2.8.8
- CHG: added 15/ 300/ 350 BPM options in Auto Heart Rate
- CHG: added 250 BPM to
Superimpose 0.1mV mains frequency noise in 201.12.1.101.15
- CHG: update the live update repository URL
Version 5.0.13.4 - 2025-04-16
- CHG: removed reference to IEC 60601-2-26 standard
Version 5.0.13.3 - 2025-03-24
- NEW: output raw data can set loop or not
Version 5.0.13.1 - 2024-11-13
- FIX: incorrect pass criteria of IEC60601-2-27 201.12.1.101.5 in Simplified Chinese version
- FIX: fail to download physionet database
Version 5.0.12.3 - 2024-05-12
- FIX: crash when outputting signals for a long time
- FIX: fail to output signals after the device is re-connected
Version 5.0.11.3 - 2024-02-15
- FIX: Assistant YY1079 test item 4.2.8.8 a) 2) - dynamic range test is not activated
Version 5.0.11.2 - 2023-09-03
- FIX: incorrect P wave amplitude in CAL05000/ CAL10000/ CAL20000
Version 5.0.10.10 - 2022-10-12
- FIX: incorrectly check tool version for live update functionality
Version 5.0.10.9 - 2022-10-06
- CHG: update assistant test description
Version 5.0.10.8 - 2022-10-05
- CHG: update assistant test description
Version 5.0.10.7 - 2022-09-26
- FIX: incorrect amplitude when variable dc is enable
Version 5.0.10.6 - 2022-09-20
- CHG: update assistant test description
Version 5.0.10.5 - 2022-07-20
- NEW: added support new assistant - GB9706.225/ GB9706.227/ YY9706.247
- FIX: (IEC60601-2-25) the initial DC offset is not configured to the device
- FIX: incorrect pacing duration when double pacing pulse is used
- FIX: (IEC60601-2-27) incorrect test procedure description in 201.12.1.101.14
Version 5.0.10.1 - 2021-08-26
- CHG: change the resolution of the pacing overshooting amplitude from 0.1mV to 0.01mV
Version 5.0.9.19 - 2020-10-12
- Refine auto pacing UI widget behavior.
Version 5.0.9.18 - 2019-02-13
- Non-ISP EEG be recognized by SECG AP
Version 5.0.9.17 - 2018-12-21
- FIX: the pacing overshooting is not changed when the pacing overshooting amplitude is updated
Version 5.0.9.16 - 2018-11-28
- Enhanced function of Auto-HR /Auto-Pacing
- SECG assistant: IEC60601-2-27 test item 201.12.1.101.1 modify. Add steps, let the user input the full scale number before starting the test
- SECG assistant: IEC60601-2-27 test item 201.12.1.101.2 modify. Add test Waveform, Triangle or Sine
- SECG assistant: IEC60601-2-27, test item 201.12.1.101.16 modify. Add steps, let the user input the full scale number before starting the test
- SECG assistant: IEC60601-2-25, 201.12.4.107.1.1.1 modify. Add Method B) C) D)
- SECG assistant: IEC60601-2-25 test item 201.12.4.108.3.2 modify. Method B: triangle pulses 1 mV, 50 ms, 120BPM
- SECG assistant: IEC60601-2-25 test item 201.12.4.109 modify. Test 1: Pacing add "Off (Vref)" (Pulse Amplitude is zero). Test 2: add "Pulse, 2 mV, width 100 ms"
- Calibration mode: Variable DC Offset Limit from 5% modify to 1%
Version 5.0.0.5 - 2016-08-09
- Solution for the inaccurate pacing duration while playing double pulse (and single pulse with 0.3/0.6 ms duration).
- Enhance SECG Assistant.
Version 5.0.0.4 - 2016-04-19
- Assistant I: Change the T-Wave amplitude in -2-27:2011, 201.12.1.101.15 from 0.2 mV to 0mV.
- Support to play decimal raw data. (e.g. Sampling Rate = 100.582 Hz.)
- Support Hot Plug
Version 5.0.0.3 - 2016-01-26
- SECG Assistant: Add new field let user enter specific testing time under the test requirement - IEC 60601-2-27, 201.12.1.101.11 Baseline reset.
- SECG Assistant: Solve the issue for "System will crash after inputing 0mV under the test requirement - IEC 60601-2-27, 201.12.1.101.17 Tall T-wave rejection capability".
Version 5.0.0.2 - 2015-11-24
- Auto Pacing: Add new functions - "Select All", "Clear All", "Backward", and "Afterward".
- Auto Pacing: Solve the issue for drawing wrong waveform after using "Backward" and "Afterward" functions.