WhaleTeq SECG5.0 Changelog
Limitation:
- only when using single pacing can the pacing duration be set to below 0.1 ms
- only support outputting ECG + Pacing (Single) + Respiration signals; double pacing is not supported
IMPORTANT:
- only supported FW version 1.21.9 and later
Version 1.5.10.12 - 2026-06-10
- NEW: Add R/IR output channel selection on the PWTT page
- CHG: R and IR can only be played one at a time when Sync Pulse is off (WAP2012, WAP2112)
- CHG: Output R and IR with the same AC/DC in asynchronous mode (WAP2012)
Version 1.5.10.11 - 2026-06-04
- FIX: signal outputting is stopped when output lead status is updated
- FIX: the parameters of auto pacing are not applied
Version 1.5.10.7 - 2026-05-16
- CHG: adjust the play data sample rate to 5K if the respiration module is connected
Version 1.5.10.6 - 2026-04-22
- FIX: failed to add noise to the playback data
Version 1.5.10.5 - 2026-04-15
- FIX: numeric controls behave incorrectly
Version 1.5.10.4 - 2026-04-09
- FIX: failed to configure apnea when playing raw data
Version 1.5.10.3 - 2026-04-01
- FIX: incorrect pacing duration if respiration module is connected
- FIX: fail to output PPG signal (WAP1001)
Version 1.5.9.33 - 2026-01-29
- NEW: Add a stop button to SpO2 page
- FIX: The channel selector on SpO2 page did not work
- FIX: The title of update external module dialog went wrong after changing language
- FIX: Parameters on the SpO2 and PWTT pages may be reset after page changing
- FIX: Wrong initial DC value for in Auto Test SpO2
Version 1.5.9.29 - 2026-01-14
- CHG: Auto scale PPG waveform to fit the max AC
- FIX: No PPG channel options in asynchronous mode
- FIX: Drop-down buttons scroll could not stop at the correct positions
- FIX: The PPG setting "AC above DC" might not work
- FIX: The PPG AC setting in "Load Raw Files" only had 2-digit integer
- FIX: The page settings in ECG and SpO2 page might not be fully applied
Version 1.5.9.28 - 2026-01-08
- FIX: Crash when showing error message dialog
Version 1.5.9.27 - 2026-01-06
- NEW: Add R-curve tables for PPG module WAP2112
- CHG: Baseline Reset Test no longer restricts signal frequency to 10 Hz
Version 1.5.9.26 - 2025-12-10
- FIX: Wrong digit of AC input in SpO2 Table page
- FIX: Display issue in Auto Test SpO2 page
Version 1.5.9.25 - 2025-12-09
- FIX: Display issue of number input widget in SpO2 Infrared group
- FIX: Display issue when changing tab during playing on Auto Test SpO2 page
Version 1.5.9.24 - 2025-12-08
- NEW: Support external module firmware update for PPG module WAP2112
- CHG: Change the min DC value for WAP2112 from 300 to 500
- FIX: Display issue of the strength bar and reference DC values in Auto Test SpO2
- FIX: Wrong text format in Output Summary
Version 1.5.9.23 - 2025-12-01
- FIX: incorrect length of HRV test pattern
- CHG: fine tune the signal strength detection (PPG module WAP2112)
- CHG: fine tune the SpO2 accuracy (PPG module WAP2112)
Version 1.5.9.21 - 2025-10-02
- CHG: update PPG DC range to 500mV ~ 3000mV (PPG module WAP2112)
- CHG: in SpO2 page, when the'Play' button is clicked, the button becomes 'Apply' button, and users can continue to edit settings
- CHG: in Auto SpO2 page, added support to configure the IR-PI value
- FIX: update the ECG signal frequency to 10Hz if the BaselineResetTest is enabled
Version 1.5.9.20 - 2025-09-16
- CHG: update PPG AC range to 0.1mV ~ 300mV (PPG module WAP2112)
- CHG: added
AC above DC setting
Version 1.5.9.19 - 2025-09-08
- CHG: added Overshoot 8 ms option in
Auto Pacing
- FIX: the ECG signal is not outputted if the BaselineResetTest is enabled
Version 1.5.9.18 - 2025-08-11
- CHG: added 30/ 300/ 350 BPM options in
Auto Heart Rate
Version 1.5.9.17 - 2025-08-07
- CHG: fine the tune the SpO2 accuracy for WAP2012
Version 1.5.9.15 - 2025-07-01
- FIX: the delay between pacing and overshooting
Version 1.5.9.12 - 2025-06-10
- FIX: in SpO2 page, AC/ DC/ PI controls in Red/ Infrared may be disabled
Version 1.5.9.11 - 2025-05-27
- CHG: fine tune the firmware update performance
- FIX: incorrect respiration information in ECG page
- FIX: incorrect ECG amplitude if modulation respiration is enabled
- CHG: moved PPG AC below DC
Version 1.5.9.8 - 2025-05-12
- CHG: update the PPG PI range to 0.01% ~ 20% (WAP2112)
Version 1.5.9.7 - 2025-05-07
- FIX: incorrect waveform type when running Auto SpO2 (WAP2112)
- FIX: waveform is incorrectly displayed in 0.5mm/mV
Version 1.5.9.6 - 2025-05-06
- CHG: added support WAP2112 PPG module
Version 1.5.9.1 - 2025-03-20
- FIX: (PWTT page) crash if BPM is set to 190
Version 1.5.8.2 - 2024-12-27
- FIX: in auto sequence crash in long-run test
- FIX: in PWTT page, the enabled status of PPG setting is not correct
Version 1.5.8.1 - 2024-11-25
- FIX: in auto sequence, data after the 10th entry is not visible
- CHG: added support to load PWTT and SpO2 file in Auto Sequence
- NEW: in SpO2 page, added support to reset to default settings, save settings to file, and load settings from file
NOTE:
- in auto sequence, up to 20 scripts can be added
Version 1.5.7.7 - 2024-08-19
- FIX: failed to update firmware
Version 1.5.7.6 - 2024-08-09
- FIX: failed to run PPG sampling
- FIX: failed to get PPG trigger level
Version 1.5.7.4 - 2024-07-30
- FIX: incorrect ECG signal when outputting ECG + Pacing + Respiration
- CHG: update the minimum value of ECG Respiration BrPM to 1
IMPORTANT:
- When outputting ECG + Pacing (Single) + Respiration, the ECG voltage and the pacing amplitude is limited to +/- 20mV
- ECG + Pacing (Double) + Respiration is not supported; the output signal will produce unexpected results
Version 1.5.7.3 - 2024-07-26
- FIX: 'White' noise is not supported
- NEW: added support to load respiration raw data (only supported when respiration module is connected)
- CHG: update the minimum value of ECG pacing duration to 0.1 ms if respiration module is connected
- FIX: "PWTT" page is not disabled if respiration module is connected
- FIX: ECG signal is incorrect for AHA txt file
- CHG: added support to configure the sample rate for Dump Raw Data function
- FIX: in ECG Special Functions > Auto page, the output lead RA is by default enabled
Version 1.5.6.49 - 2024-06-19
- FIX: a delay appears when outputting pwtt signals in the beginning
- FIX: change the amplitude limitation of dynamic range test to 40mV
- FIX: fail to enable the save wave add-on button after stopping waveform signal output
- FIX: fail to run frequency scan function
- FIX: the prompt message is missing when running standard assistant
- FIX: the pacing signal is not outputted in standard assistant
- FIX: update assistant pass criteria
- FIX: pacing should be 30ms before the P wave
Version 1.5.6.31 - 2024-04-09
- CHG: disable electrode V1~V6 if respiration module is connected
- FIX: the time range in Auto PWTT setting form is incorrect
- FIX: the pacing amplitude adjustment is inappropriate
Version 1.5.6.29 - 2024-03-27
- CHG: change the decimal places of QRS Amplitude to 2
- FIX: incorrect ECG signal amplitude if respiration module is connected
Version 1.5.6.28 - 2024-03-19
- CHG: update "Dump Raw Data"; only save AC data
Version 1.5.6.27 - 2024-03-11
- FIX: incorrect pacing output after respiration apnea
- NEW: added support respiration module (only supported FW version 1.3.31.04 and later)
Version 1.5.6.26 - 2024-03-04
- FIX: incorrect pacing duration
Version 1.5.6.25 - 2024-02-23
- CHG: update "Dump Raw Data"; update to fixed sample rate 1K and only save one cycle data
Version 1.5.6.23 - 2024-01-26
- CHG: update the maximum respiration BrPM to 200
- CHG: added support ECG noise frequency - 100/ 120/ 150/ 180/ 200/ 240 Hz
- FIX: incorrect show the error message after successfully enable assistants
- FIX: the variable dc is not restored to 0mV after stopping signal output
- CHG: in Auto SpO2 page, show the stable DC value after stopping signal output
Version 1.5.6.20 - 2024-01-17
- NEW: added Auto SpO2 page
Version 1.5.6.2 - 2023-11-03
- FIX: Redundant option "0" in Inhale : Exhale Ratio
- CHG: update the range of pacing duration to 0.03ms to 3ms (only supported FW version 1.3.22.12 and later)
IMPORTANT:
- only supported FW version 1.3.22.12 and later
Version 1.5.5.19 - 2023-10-04
Note: Please use firmware version 1.3.19.02+ to ensure that all changes work.
Common
- NEW: Add function to clear the cache and restart the app
- CHG: Improve file browse dialog, add automatic scrolling
- CHG: Improve the performance of reading variable DC calibration data from SECG
- FIX: Variable DC calibration value may be incorrect when the connected SECG serial number changed
- FIX: The dialog box message does not reset when updating FW consecutively
- FIX: App crashes during version updates
- FIX: App resources do not restore when downgrading versions
- FIX: App may crash when accessing the file names with non-ASCII characters
ECG
- CHG: The chart scale of ECG page now can restore when switch back from PPG pages
- CHG: Adjusted Auto Sequence: If it includes deleted scripts, but there are other valid scripts, it can still be executed. Invalid ones will be automatically skipped.
- FIX: Status not immediately update issue when deleting scripts in Auto Sequence
- FIX: The app may crash when reading an old script
- FIX: The app may crash when reading a txt file which format is incorrect
- FIX: The app may crash when select a .hea file in File Replay
- FIX: Incorrect 620kΩ status in Output Summary
- FIX: Incorrect BPM and QRS Duration values in Output Summary when running Auto Heart Rate
- FIX: The graph may not be cleared when switching from ECG page to other pages
- FIX: ECG Respiration apnea time 60s setting does not work
- FIX: Incorrect waveform amplitude in UI when enable ECG Respiration
- FIX: Incorrect P wave amplitude in CAL waveform
- FIX: Rounding error of BPM in decimal places
PPG
- NEW: Add support for PPG reflective module and green light module
- NEW: Add buttons to browse Raw Files
- NEW: Add buttons to set Sampling, and move them to left side
- CHG: Play Raw Files now supports Sync Pulse and Inverted Pulse
- CHG: Updated method for PPG Play Raw second setting Output Setting, consistent with AECG
- CHG: Improve the performance of PPG Sampling waveform
- CHG: Trigger Level setting is now updated when clicking OK, except for restoring to default values, which are set immediately
- CHG: Update Sync Pulse mode, now all channels output the signal respectively instead of outputting single waveform
- FIX: Improve performance when output PPG first time from start-up
- FIX: Correct the definition of PPG inverted waveform (inverted refers to output being opposite of the original waveform, but it does not reverse in the UI), and set inverted as default
- FIX: PPG module output may be abnormal after using Sampling
- FIX: Incorrect PPG waveform amplitude when using Respiration Baseline mode or Raw Files
- FIX: Signals not reset when changing files in PPG Load Raw File page
Spo2
- NEW: Add function to import and delete SpO2 Table
- NEW: Add new SpO2 Tables: P_R-curve(T), G_R-curve(T)
- CHG: Update the default SpO2 Table file names, please remove the old copy manually
- FIX: Incorrect initial values of PI and SpO2
- FIX: Incorrect behavior when adjusting formula coefficient in SpO2 (should recalculate SpO2)
- FIX: Too low accuracy of R Value decimal, only three places
PWTT
- NEW: Add new function: Auto PWTT
- NEW: Add functions to save, load, and reset parameters
- NEW: Add Sampling in this page
- NEW: Add Load Raw Files function
- CHG: Now all kinds of PPG waveforms can be loaded from the SpO2 page
- CHG: When PWTT page loads SpO2 parameters, parameters can be adjusted using SpO2 value
- FIX: PI value may be correct when loading parameters from SpO2 page
- FIX: Remove redundant parameters when loading ECG 2-27 waveform (only three items left now)
Standard Assistant
- FIX: The buttons on parameters dialog may be covered
Version 1.5.4.9-hotfix - 2023-07-04
- Fix that waveform does not show when running frequency scan
Version 1.5.4.7 - 2023-07-04
Changed:
- Release the Standard Assistant files
- GB9706.225-2021 (2023/6/30)
- GB9706.227-2021 (2023/6/30)
- YY9706.247-2021 (2023/6/26)
Fixed:
- Fix the crash in File Replay when playing the files which the file name are too long.
Version 1.5.3.0 - 2023-05-17
New:
- Support PPG-2TF-660 module, and add pages "SpO??quot; and "PWTT"
Note: Require firmware version 1.3+
- Add new Standard Assistant GB9706.225-2021?GB9706.227-2021?YY9706.247-2021 (beta)
- Add serial number of connected PPG module on title bar
- Add the 3rd, 4th code of firmware version on Setting dialog (1.3+)
- Add scroll bar on Output Summary box
Changed:
- Change the page title "Signal Waveform" to "ECG"
- Ignore file extension case when searching update files
- Improve the performance of upgrading firmware
- Improve the preview waveform of Baseline Overload Test, and not reset when turning off
Fixed:
- Fix the display issue of Hardware ID
- Fix the issue of automatically adjusting ECG parameters
- Fix the crash when saving SDK Log without a USB device
- Fix the output and preview waveform of Baseline Overload Test
- Fix some spelling mistakes of Output Summary
- Fix the HRV parameters display issue in Output Summary
- Fix the issue of loading DC calibration data
- Fix the ECG wave modulation of Respiration, and make it synchronous with PPG wave modulation
- Fix the display issue of ECG Noise 50Hz
- Fix the preview waveform of Pacing Overshooting
- Fix the failure of loading some txt files
- Fix the display issue of the file name with non-alphabetic characters
- Fix the issue of reconnecting to hardware
- Fix the failure of importing license
- Fix missing description line in exported raw data file
- Fix missing waveform when outputting pacing
- Fix the crash when saving files to USB device in some case
- Fix the input variable dialog size mistake of Standard Assistant
- Fix the combo box display issue of Standard Assistant
Version 1.4.5.5 - 2022-12-16
New:
- Add feature of DC offset variable mode, and make it available in new scripts
- Add firmware version showing on Setting dialog
- Add feature of restoring the last status of Auto Sequence
Changed:
- Update the upper limit of script playing period of Auto Sequence, and change the UI widgets
- Update the feature of ECG Respiration wave modulation that the 3 modulation can be mixed now, and update the script
Note: the wave modulation setting in previous scripts will be no effect anymore.
- Update T wave in ECG waveform: fixed period 180ms, and will be removed when it exceeds the QT Interval
Fixed:
- Fix missing error message when failed to copy files to USB device
- Fix that there will be one extra directory when exporting SDK Log
- Fix that file names which is too long cannot be displayed on Auto Sequence
- Fix that the parameters may not be adjusted when the ECG frequency exceeds 2Hz
- Fix the display issue of ECG Respiration parameters
- Fix that the parameters of ECG Respiration wave modulation are sometimes no effect
- Fix that the parameters of Baseline Overload Test are sometimes no effect
- Fix the display issue after turning off the screen when playing
- Correct the Pacing waveform start position of non-ECG waveforms
- Correct the Pacing Overshooting waveform
- Fix that the Common Mode switch cannot be used after playing
- Fix that Common Mode would not be toggled immediately
- Fix some issue of Standard Assistant IEC60601-2-27:2011
Version 1.3.3.11 - 2022-03-22
- ECG will be set to IEC60601-2-27 waveform when pacing is turned on.
- Improve the performance of Auto Sequence.
- Fix pacing waveform pulse time issue.
- Fix pacing overshooting waveform.
- Fix Assistant IEC60601-2-27 bugs
- Fix Assistant YY1139 bugs
Version 1.3.2.2 - 2022-01-07
- Add new Standard Assistant: YY1139-2013
- Fix crash when switching IEC60601-2-27 HRV Test patterns
- Fix failing to return to the main window after overwriting the dump raw file
- Fix dumping raw data no device issue.
- Fix crash if a file name is too long in File Replay.
- Fix several mistakes in Output Summary
- Fix the noise of pacing waveform of specific parameters.
Version 1.3.1.11 - 2021-11-29
- Improve the sample rate to 20kHz
- Improve the performance of waveform graph
- Synchronize electrode button when frequency scan
- Fix the bug of BPM button
- Fix the bug of setting dialog displaying
- Fix the pacing waveform with overshooting
- Fix Auto Heart Rate output summary
- Fix AAMI EC13 preview waveform
- Fix the waveform of ECG frequency scan
- Fix pulse width button state
- Fix time label display bug
- Fix Assistant IEC60601-2-25 bugs
- Fix Assistant IEC60601-2-27 bugs
- Fix Assistant YY1079-2008 bugs
- 4.1.2.1 electrode display bug
- 4.1.4 several bugs
- 4.2.6 output waveform bug
- 4.2.7.3 electrode display bug
- 4.2.8.1 electrode display bug
- 4.2.8.8 specific item no output issue
- Fix crash when switching assistant tests
Version 1.2.1.23 - 2021-09-08
- Update Assistant YY0885-2013
Version 1.2.1.21 - 2021-08-25
- Change the step of pacing overshoot amplitude from 0.1 to 0.01 mV.
Version 1.2.1.20 - 2021-08-05
Common:
- Change the default values of some items in Auto Heart Rete
- Fix short duration pacing waveform (make it more accurate)
- Fix negative pacing amplitude overshooting
- Fix Auto Pacing bugs
- Fix Baseline overload test bugs
- Fix the output summary of Auto Pacing
Assistant:
- Add pre-test command SetDCOffsetVariableMode(false)
- Change IEC60601-2-27 Global Parameter: Full scale output value range to 0~10
- Modify YY1079-2008 4.2.8.4
- Fix 201.12.4.107.2 (IEC60601-2-25:2011)
- Fix 201.12.1.101.8 (IEC60601-2-27:2011)
- Fix 201.12.1.101.9 (IEC60601-2-27:2011)
- Fix 201.12.1.101.12 (IEC60601-2-27:2011)
- Fix 201.12.1.101.13 (IEC60601-2-27:2011)
- Fix 201.12.1.101.17 (IEC60601-2-27:2011)
- Fix Global Parameter radio button text bug (IEC60601-2-27:2011)
- Fix 51.104.3 (YY0782-2010)
- Fix 4.1.4 overshoot amplitude spec (YY1079-2008)
- Fix 4.2.8.1 DC output (YY1079-2008)
- Fix 4.2.8.1 input variable (YY1079-2008)
- Fix 4.2.8.8 (YY1079-2008)
Version 1.2.1.1 - 2021-06-02
Common:
- Reduce APK size
- Modified the disabled-color of switch buttons
- Add loop option on standard waveform and file replay page
- Add delete button in file browse dialog
- Show warning before overwriting file
- Disable the script in sequence which file is lost
- Implement Sine Wave dynamic range test
- Implement standard waveform raw data dumping
- Fix ECG parameters reset and auto-set issue
- Fix NST Noise on HRV pattern issue
Assistant:
- Support "?(Please fill the ref value)" for unsolved reference value.
- Support multiple report value input.
- Support noise on ANE20000 waveform
- Fix the side effect of multiple input result table value.
- Fix YY Standard, wrong table header text shown.
- Fix IEC60601-2-27 bugs
- Fix IEC60601-2-25 bugs
- Fix YY1079 bugs
- Fix YY0782 bugs
Version 1.1.23.6 - 2021-02-25
- Fix bugs in Assistant IEC60601-2-27
- Fix ECG parameters auto adjustment
- Implement electrode button state in Assistant
- Implement noise on ANE20000 waveform
Version 1.1.23.1 - 2021-01-28
- Fix pacing of some Assistant test items
- Reduce the size of APK file
Version 1.1.22.17 - 2021-01-27
- Improve USB driver
- Change the minimum pulse width to 1 ms
- Update Assistant files
- Fix the potential bug of "dump standard"
- Fix the display bug of baseline overload test
- Fix standard waveform IEC60601-2-47 pattern 1
- Fix auto adjusting for ECG parameters
- Fix several bugs of Assistant
- Fix the display bug of "Auto Sequence"
Version 1.1.22.2 - 2020-11-04
- Initial version