WhaleTeq AECG100 Changelog
Limitation:
- Auto Sequence page: support PPG noise frequency only upto 1K Hz
- only supported FW version 1.33.9.22 and later
Version 1.0.10.42 - 2026-01-07
- FIX: prevent from updating the incorrect MCU image (PPG module WAE2112)
Version 1.0.10.41 - 2026-01-05
- CHG: added R-curve tables for PPG module WAP2112
Version 1.0.10.40 - 2025-12-24
- FIX: incorrect PPG noise amplitude
Version 1.0.10.39 - 2025-12-20
- FIX: PPG waveform settings are not corrected saved into device mode settings
- FIX: in Auto Sequence page, noise amplitude of ECG waveform is incorrectly outputted
Version 1.0.10.38 - 2025-12-17
- FIX: PPG invert/sync settings are incorrect after doing Recall
- FIX: UI hangs if clicking PPG setting button during playing data
Version 1.0.10.36 - 2025-12-10
- FIX: crash when running PPG database (WAP2112)
Version 1.0.10.35 - 2025-12-06
- FIX: incorrect sampling data (WAP2112)
- FIX: in PPG page, the channel selection is invisible after changing PPG modules
- FIX: incorrect ECG pacing duration
- FIX: added support MCU live update for PPG module WAP2112
Version 1.0.10.28 - 2025-11-06
- FIX: in Auto SpO2 page, R-AC suggested value is not given in a new empty line in SpO2 Table
- CHG: fine tune the signal strength detection (PPG module WAP2112)
- CHG: fine tune the SpO2 accuracy (PPG module WAP2112)
Version 1.0.10.25 - 2025-10-27
- CHG: update PPG DC range to 300mV ~ 3000mV (PPG module WAP2112)
Version 1.0.10.23 - 2025-10-08
- CHG: fine tune to improve frequency accuracy
Version 1.0.10.22 - 2025-10-01
- CHG: in Auto SpO2/ SpO2/ PPG pages, when 'Signal' > 'Play' button is clicked, the button becomes 'Apply' button, and users can continue to edit settings
- FIX: in Auto Sequence page, incorrectly configure the PPG module (PPG module WAP2112)
Version 1.0.10.20 - 2025-09-23
- FIX: could not download mimic db
- CHG: update PPG DC range to 500mV ~ 3000mV (PPG module WAP2112)
Version 1.0.10.18 - 2025-09-18
- FIX: incorrect PTTp time when running in standalone mode
- FIX: in PWTT page, crash when running Database Player
- FIX: incorrectly display FW build version of the loaded FW image
Version 1.0.10.16 - 2025-09-09
- CHG: update PPG AC range to 0.1mV ~ 300mV (PPG module WAP2112)
- CHG: in PPG page, remove AC Offset setting
Version 1.0.10.14 - 2025-08-11
- FIX: in PPG page, DC value may not correctly updated in some cases
- CHG: update the DC minimum value based on AC (if AC is below DC)
- FIX: in PPG page, crash when adjusting SP voltage
- CHG: update the live update repository URL
- CHG: save the setting of PPG AC signal is above or below DC
IMPORTANT:
- By default, PPG AC is below DC. Therefore, from this version onward, the measured SpO2 values may differ slightly from previous versions.
Version 1.0.10.8 - 2025-05-20
- CHG: update the increment value of PPG DC from 10 to 100
- CHG: in Auto SpO2 page, added support to configure the IR-PI value
Version 1.0.10.7 - 2025-05-09
- CHG: update the PPG PI range to 0.01% ~ 20% and PPG AC range to 0.1mV ~ 60mV (PPG module WAP2112)
- CHG: added support to configure PPG AC signal is above or below DC
- FIX: transmission rate is not correctly configured for PPG module WAP2112
Version 1.0.10.5 - 2025-04-11
- CHG: moved PPG AC below DC
Version 1.0.9.4 - 2025-01-21
- FIX: in PPG page, crash when saving waveform raw data if PPG noise is not Off
Version 1.0.9.1 - 2025-01-06
- CHG: the respiration algorithm can be multi selected
- CHG: added support apnea in ECG and PPG respiration algorithm
- FIX: crash when outputting signal by play data mode with noise amplitude 0mV
- FIX: fail to download physionet database
Important:
- If the saved waveform configuration file includes the respiration function, this configuration file will not be compatible with the current version.
Version 1.0.8.19 - 2024-10-30
- FIX: DC is incorrectly adjusted (PPG module WAP2012)
Version 1.0.8.16 - 2024-08-31
- FIX: the noise signal is not added to PPG in the simulation waveform
Version 1.0.8.15 - 2024-08-27
- FIX: AC/ DC values are incorrect in Player > Play for PPG module WAP2012
- FIX: in Auto Sequence page, hang issue occurs if the output status of the loaded PPG file is not enabled
- FIX: in PPG page, noise is not added if the respiration is enabled
Version 1.0.8.14 - 2024-08-19
- NEW: added PPG noise frequency 25 Hz (only supported FW version 1.14.44.19 and later)
Version 1.0.8.12 - 2024-06-27
- FIX: in Auto Sequence page, ECG pacing (.ecg) is invalid (the pacing duration minimum value is limited to 0.1 ms)
- NEW: added PPG noise frequency 0.5/ 1/ 2/ 3/ 5/ 10 Hz (only supported FW version 1.14.42.1 and later)
- FIX: in Auto Sequence page, crash if loading the file saved from ECG page with sine/ 50Hz waveform
Version 1.0.8.10 - 2024-03-20
- FIX: in Auto Sequence page, failed to load the PPG setting file (*.ppg) (only for PPG module WAP2009/ WAP2010/ WAP2013/ WAP2014)
- CHG: in SpO2 and PPG page, if the settings are saved as a PC file, the output enabled status are also saved
- FIX: in PPG page, the respiration is not effective if the PPG signal only outputs IR
Version 1.0.8.8 - 2024-02-15
- NEW: added support new PPG module WAP2013 (PPG-2RS-880)/ WAP2014 (PPG-2RS-940)
Version 1.0.8.7 - 2023-09-15
- NEW: added support new PPG module WAP2009 (PPG-2R-880)
Version 1.0.8.4 - 2023-08-25
- CHG: update the range of pacing duration to 0.03ms to 2ms (only supported FW version 1.15 and later)
- FIX: save incorrect PWTT raw data
Version 1.0.7.3 - 2023-06-16
- NEW: added support YY 9706.247 (2021)
- FIX: fix ui defects
- FIX: incorrect ECG signal if PWTT respiration is enabled
Version 1.0.6.86 - 2023-02-22
- CHG: change the minimum value of PPG DC to 30mV (PPG module WAP2012)
- CHG: update the warning text in Auto SpO2 page (PPG module WAP2012)
- FIX: failed to popup warning message if the DUT is not detected (PPG module WAP2012)
- FIX: in some cases, failed to get the correct R value in SpO2 table (PPG module WAP2012)
- CHG: update the notification text when outputting in SpO2 table dialog
Version 1.0.6.80 - 2022-09-27
- CHG: in Auto Test SpO2 page, fine tune the final stable DC (PPG module WAP2012)
Version 1.0.6.79 - 2022-08-23
- FIX: the PI of DUT is not accurate when testing in auto sequence (PPG module WAP2012)
Version 1.0.6.78 - 2022-07-28
- CHG: in Auto Test SpO2 page, add warning message about the SpO2 accuracy
Version 1.0.6.76 - 2022-07-08
- CHG: adjust DC levels to make PI more accurate (PPG module WAP2012)
Version 1.0.6.73 - 2022-06-13
- FIX: in FPGA programming mode, the device may block and fail to return to normal state
- FIX: in ECG page, the time constant is incorrect for Exponential waveform
- FIX: failed to copy correct values in SpO2 Table dialog
- FIX: in PPG page, the supported channels are incorrect
Version 1.0.6.71 - 2022-06-10
- FIX: in PWTT page, the measured PTTp becomes incorrect if BPM > 125
- CHG: in Auto Test SpO2 page, update the DC levels for the transmission rates
- CHG: update the names of SpO2 sample tables
Version 1.0.6.67 - 2022-05-26
- CHG: if PPG module WAP2012 is connected, show the ambient light mode in the simulation graph
- CHG: update transmission rate options (PPG module WAP2012)
Version 1.0.6.63 - 2022-04-29
- CHG: in ECG page, support frequency range to 0.05Hz ~ 5Hz for Rectangle Pulse/ Triangle Pulse
- CHG: in ECG page, set the baseline of Rectangle Pulse/ Triangle Pulse/ Exponential waveforms to zero voltage
Version 1.0.6.59 - 2022-04-20
- FIX: crash when running auto sequence items
Version 1.0.6.57 - 2022-04-13
- FIX: (PPG module WAP2012 connected) in PWTT page, if 'Sync Pulse' is off, output the selected channel and display the selected in the PPG checkbox
Version 1.0.6.56 - 2022-04-08
- FIX: the fixed DC offset is not configured to device when running the sequence items in Auto Sequence page
- CHG: by default, set the Transmission Rate to Medium Finger in Auto SpO2 page
Version 1.0.6.54 - 2022-03-25
- FIX: crash when detecting the signal strength
- FIX: crash when restoring PPG waveform from standalone mode settings
- CHG: move 'Ambient Light' setting to PPG context menu
Version 1.0.6.52 - 2022-03-18
- CHG: remove the Signal Strength function under PPG button
Version 1.0.6.51 - 2022-03-16
- CHG: by default, turn off PPG Dynamic LED Strength feature
- CHG: support to output only R or IR for PPG module WAP2012 if SynPulse is OFF
- NEW: added support to auto find the trigger level (only available in PPG module WAP2012; run in Auto SpO2 page)
Version 1.0.6.49 - 2022-03-08
- CHG: removed support to adjust PPG LED Level
- CHG: remove waveform type in Auto SpO2 page
- NEW: added support new PPG module WAP2008
Version 1.0.6.48 - 2022-02-25
- CHG: by default, do not turn off PPG Dynamic LED Strength feature
Version 1.0.6.47 - 2022-02-17
- CHG: support to configure transmission rate for PPG module WAP2012
Version 1.0.6.46 - 2022-01-07
- FIX: failed to output PPG waveforms; not compatible with old fw
- FIX: in PWTT page, the PPG waveform invert status is incorrect for PPG module WAP2007
- FIX: in PWTT page, it's failed to run respiration
- FIX: fail to output ECG waveform when noise is enabled
Version 1.0.6.40 - 2021-12-23
- FIX: incorrect LED output for PPG module WAP2007; LED-R should not outputted if LED-R is not included in pulse group setting
Version 1.0.6.39 - 2021-12-21
- CHG: update the model description of PPG module WAP2007 to R(660nm)/ IR(880nm)
- CHG: remove Player tab in new SpO2 page for PPG module WAP2012 (still in development)
- CHG: show the FW build version in About dialog
- CHG: update Pulse Configuration; removed one-trigger mode functionality; added IR-First functionality
- CHG: keep the original SpO2 page when PPG module WAP2012 is connected
Version 1.0.6.36 - 2021-12-16
- CHG: update PPG module name; WAP2012: PPG-2TF-660
- CHG: removed support PPG module WAP2014
- CHG: fine tune FPGA update process
- CHG: support 16b-bit DAC for PPG module WAP2012
- NEW: support to configure the ambient light for PPG module WAP2012 (still in development)
Version 1.0.6.33 - 2021-11-16
- FIX: for FW compatible, when 3-channel simulation PPG module is connected, set all the output enable bits on
Version 1.0.6.32 - 2021-09-07
- FIX: in PPG page, exception issue may occur when adjusting waveform voltage
Version 1.0.6.31 - 2021-09-03
- NEW: in PWTT page, added support to lock PTTp or PTTf when BPM is changed
Version 1.0.6.30 - 2021-08-24
- CHG: hide "Dynamic LED Strength Adjustment" feature (in development)
- FIX: incorrect RLD unit
Version 1.0.6.28 - 2021-08-13
- CHG: check the FPGA hardware status; it it's not activated, don't show the PPG information
Version 1.0.6.25 - 2021-08-11
- FIX: incorrect ECG exponential waveform; the baseline should be shifted
Version 1.0.6.24 - 2021-08-09
- FIX: required to stop PD sampling when detecting DUT signal strength
- FIX: only allow only one Signal Strength window being opened
- FIX: failed to only output PPG Ch2 waveform
Version 1.0.6.22 - 2021-08-06
- FIX: FGPA calibration data is incorrectly cleared when restoring factory setting
Version 1.0.6.21 - 2021-07-16
- FIX: refer to incorrect PD mapping table in the output setting of PPG play raw
- FIX: can not configure the PD mapping table for PPG Ch3
Version 1.0.6.20 - 2021-07-08
- FIX: in PWTT page, incorrectly output PPG Ch2 AC
- FIX: in PPG page, incorrectly display the outputted waveform
Version 1.0.6.19 - 2021-06-28
- NEW: added support LED output for PPG module WAP3001
Version 1.0.6.18 - 2021-06-24
- FIX: IR should be outputted for PPG module WAP2007
- CHG: added support PD sampling for PPG module WAP2012
- NEW: added support to configure Ch3 LED/Trigger level for PPG module WAP3001
Version 1.0.6.16 - 2021-06-09
- FIX: the respiration settings of ECG and PPG are not saved to and restored back from the file or the device
- NEW: added support new PPG module WAP3001
Version 1.0.6.15 - 2021-06-07
- FIX: in Auto Sequence page, PWTT modulation respiration not works
Version 1.0.6.14 - 2021-05-31
- CHG: in PPG page, still keep the PPG respiration setting when the respiration is disabled
- NEW: in PWTT page, added support modulation respiration
- NEW: added support new PPG module WAP2007
Version 1.0.6.11 - 2021-03-26
- FIX: the PWTT preview graph incorrectly display the time indicator of PTTp/PTTf
- FIX: in SpO2 page, if R PPG module is connected, the enabled status of the Signal > Output button is not correctly updated
Version 1.0.6.9 - 2021-03-18
- FIX: incorrect ECG exponential waveform; the pulse width is not necessary
Version 1.0.6.7 - 2021-03-02
- FIX: the working mode is not correctly configured for single-LED PPG module
Version 1.0.6.6 - 2021-02-05
- NEW: added support Dynamic LED Strength Adjustment function
Version 1.0.6.5 - 2021-02-01
- FIX: potential blocking issue when playing raw data
Version 1.0.6.4 - 2021-01-27
- NEW: added support new PPG module WAP2012/ WAP2014
- NEW: added support Finger Signal function for PPG module WAP2012/ WAP2014
- FIX: when R PPG module is connected, the enabled PPG channgles are not correctly updated based on Pulse Configuration
- FIX: in PPG page, when R PPG module is connected, the BPM is not all synchronized to all channels
Version 1.0.6.1 - 2020-12-01
- NEW: support to save the items in SpO2 Table dialog to *.seq file and import to Auto Sequence page
- NEW: in Auto Sequence page, if Shift key is pressed, the duration of all items will be updated synchronously
- FIX: blocking issue in SpO2 page when doing PD sampling and creating SpO2 table in the same time
- FIX: if the PPG is other than PPG waveform, the amplitude in the simulation graph is incorrect
Version 1.0.5.34 - 2020-08-25
- PPG page: support PPG respiration; baseline/ amplitude/ frequency modulation
- remove support new PPG module (only R LED)
- fix: the output setting of PPG in the Waveform Player is not correctly displayed
- PPG page: if the signal is outputting with the PPG respiration, the PD sampling keeps missing packets; update to do one thing in the same time
- fix: the zero voltage of ECG variable DC is not precisely configured
- PPG page: the AC offset is not saved to the device
Version 1.0.5.26 - 2020-06-15
- PPG page: temporarily disable the PPG respiration
Version 1.0.5.25 - 2020-06-10
- sync the curve colors in the preview graph with the colors in the signal output graph
- update the PPG-G default settings for R PPG module
- in signal/ player graphs, only display the enabled output channels
- fix: when R PPG module is connected, in PPG Respiration dialog, Amplitude-G is not sync with Amplitude-R
- SpO2 page: when "Use SpO2 Table" is checked, by default, select an empty Table item
- ECG page: remove ECG respiration Basic Level 500 ohm
- fix Assistant IEC 60601 2-47 201.12.4.4.101: the frequency is incorrect for digital test
- fix: when R PPG module is connected, the settings in the PPG Configuration dialog are incorrectly changed
- fix: PPG channel-1 switch sampling may be influcened by PPG channel-2 trigger status
- fix: Assistant: modify "201.12.4.4.108 Frequency response b) Triangular Pulse" to "201.12.4.4.108 Frequency response b) Sinusoidal Signal"
- SpO2 page: fix DC + DC Offset is over 3000 mV when changing DC value
- PWTT page: by default, check the 'Lock DC' checkbox instead of 'Lock AC'
Version 1.0.5.21 - 2020-04-09
- SpO2 page: by default, check the 'Lock DC' checkbox instead of 'Lock AC'
- PPG page: by default, check the 'Lock DC' checkbox instead of 'Lock AC'
- fix bug: PPG waveforms not output if Tab key is entered in SpO2 table generation dialog
- added support to save noise to waveform raw data in WhaleTeq format (license key is required)
- PPG page: remove 'Frequency Scan' function
- if PPG module R/IR is connected, popup message to notify users to connect the Aux power
- ECG page: update the range of frequency scan to 0.05~500Hz; the start frequency must be less or equal to the stop frequency
- After the waveform outputting is stopped, there are unexpected short signals are generated
- fix the Pulse Sequence in the PPG Pulse Configuration form may contain empty selection
Version 1.0.5.17 - 2020-02-03
- Auto Sequence page: fix ECG/ PPG noise not working (limitation: support PPG noise frequency only upto 1K Hz)
Version 1.0.5.16 - 2019-12-18
- support new PPG module (only R LED); if the new R PPG module is connected, the Switch sampling function will be disabled
- support to save/recall the waveform player setting to/from a file (*.pecg, *.pspo, *.ppwv, *.pppg); the player setting file can be added to auto sequence test
- added options of PTTp and PTTf in Auto PWTT testing
Version 1.0.5.4 - 2019-07-10
- PPG page: add Lock AC and Lock DC
- ECG page: add noise frequency options - 100Hz/ 120Hz
- PPG page: add noise frequency optinos - 100Hz/ 120Hz/ White Noise
Version 1.0.5.3 - 2019-07-08
- support PPG respiration to PPG-G
Version 1.0.4.61 - 2020-03-16
- fix signal not outputting in standalone mode after saving waveforms to Mode A/B/C
- fix PTTp/PTTf is not correctly configured when saving waveforms to Mode A/B/C
Version 1.0.4.58 - 2020-01-07
- fix crash when right-click in the spo2 table generation dialog if the table is empty
- fix download database failed
- fix incorrect QT interval ragne in ECG page
- fix load *.ppg files to auto sequence failed if DC > 2500mV
- fix load *.ppg files to auto sequence failed if noise frequency is 5K Hz
- fix crash when adjusting amplitude in ECG page
- fix ECG amplitude can not be adjusted to 1.00mV
- fix PPG AC Offset is incorrectly configured
- fix PPG frequency scan not working in some conditions
- PWTT page: fix incorrect PTTp time for some BPM
- PWTT page: fix exception issue when adjusting BPM
Version 1.0.4.47 - 2019-09-06
- fix crash when loading PPG waveform file (10 BPM)
Version 1.0.4.45 - 2019-08-08
- fix crash when recalling the files which are saved in the previous version2. PPG page: incorrect AC offset value is applied if the waveform type is not PPG waveform
- ECG page: support pacing duration up to 0.1 ms
Version 1.0.4.41 - 2019-07-23
- fix ECG play raw data failed in long-run test
- fix incorrect ECG signal ratio if respiration is enabled
- update ECG respiration spec: respiration rate 4 ~ 200
- SpO2 page: when adjusting AC/DC or the formula parameters, if the SpO2 value is not in the range 0-100, display the SpO2 value in red color
Version 1.0.4.18 - 2019-03-15
- PWTT page: fix 'Reset' function; the PTTp in Time Difference should be reseted to 500ms
- SpO2 Table form: immediately play the PPG waveforms based on the current selected row settings if the settings are valid; if not valid, stop playing
- PPG page: fix crash when dragging DP point in preview graph
- PPG page: fix the maximum value of AC Offset is not updated when SP is changed
Version 1.0.4.17 - 2019-03-07
- PPG page: the voltage proportional adjustment is incorrectly after Reset function
- PPG page: added support AC offset function (AC Offset maximum value = 30mV - noise amplitude - VolSP amplitude)
- fix PD/switch sampling synchronization issue
- PWTT page: add AF database in the WhaleTeq Database if it's trial version5. support to minimum 10 BPM in ECG/ PWTT/ SpO2/ PPG pages
- update the QT interval from 260ms-450ms to minimum=100ms - maximum=900ms
- disable 'Play' button in Auto Sequence page when the total time is 0 sec
- fix crash if the parameters of the SpO2 formula are invalid values
- update hardware versionnaming rule
Version 1.0.4.11 - 2019-01-18
- fix bug: in PPG page, save Waveform=Square to file, and then recall it, the waveform type is incorrectly restored to PPG
- PPG page: enhance PPG waveform voltage adjustment process; if the SP voltage is changed, the DN/DP voltage will also be proportional changed
- PPG page: when the Waveform type - Square is selected, the voltage trackbar is incorrectly displayed (only AC should be displayed)
- PWTT page: fix compatibility issue; the *.pwv file saved from 1.0.3.x vesion is not correctly restored back
Version 1.0.4.9 - 2019-01-08
- adjust the maximum value of pulse width based on the frequency in ECG page
Version 1.0.4.8 - 2018-12-26
- added support to save waveform raw data in WhaleTeq format (license key is required)
- added support PPG frequency scan
- added support to set respiration apnea cycle
- added support PPG square waveform
- in PWTT page, added support to load ECG square waveform from ECG page and load PPG square waveform from PPG page
- added support Auto PWTT testing in PWTT page
- fix when only sampling is running, it should be allowed to modify PPG settings (Invert/ Synchronized Pulse)
- the PPG Invert and Synchronized Pulse settings of the PPG items in the auto sequence are refer to the global PPG settings
Version 1.0.3.70- 2019-02-21
- fix run assistant crash
- add ECG respiration Basic Level 500 ohm
Version 1.0.3.69- 2019-01-19
- fix bug: fix incorrect max/min value of numeric control in ECG and PWTT file in Auto Sequence page
- fix PPG max amplitude is still 28mV after the noise amplitude is changed from 2mV to 0mV
Version 1.0.3.67- 2019-01-10
- update the spec of SpO2 PI range
- disable ECG respiration when ECG signal play is stopped
- fix incorrect sampling status between PPG and SpO2 pages
- fix failed to recall ppg file in auto sequence
Version 1.0.3.64- 2019-01-03
- fix WhaleTeq-format file parser if newline is inserted
- fix PD mapping table list is not immediately updated
- fix memory leak
- in PWTT page, added support Signal -> Save even when the ECG/PPG waveforms are loaded from other tabs
- fix pacing rate not synchronized when 'Sync with main function' is enabled
- fix recall failed if PPG module (R/IR) is connected
- enhance the recall performance in Auto Sequence
- check fw compatibility; show error message and close the application if the connected device fw is ver.1.4
- fix live update issues
Version 1.0.3.52- 2018-11-29
- fix crash when loading incorrect WhaleTeq-format file in play raw function
- fix incorrect enabled status of pacing setting group in pacing dialog
- disable the PPG Invert and Synchronized Pulse settings if the device is outputting
Version 1.0.3.50- 2018-11-28
- fix incorrect SpO2 preview when AC value is changed
Version 1.0.3.49- 2018-11-26
- fix calibration curve calculation issue
- fix standard assistant issues
- fix failed to play mimic db
Version 1.0.3.46- 2018-11-19
- fix incorrect pacing rate in standard assistant
Version 1.0.3.45- 2018-11-16
- fix ui issues in SpO2 page
- adjust time periods of PPG wavform based on the time ratio when BPM or frequency is changed
Version 1.0.3.44- 2018-11-16
- fix play raw data issue; non-looping mode is not workable
- fix failed to play pwtt in auto sequence
Version 1.0.3.43- 2018-11-13
- added support to choose which channels to output in PPG page
- fix spo2 table generation issues
- fix incorrect PTTp calculation
Version 1.0.3.39- 2018-10-26
- fix crash issue: incorrectly-defined spo2 table may results in crash
Version 1.0.3.38- 2018-10-25
- SpO2: added support polynomial calibration curve
Version 1.0.3.35- 2018-09-13
- update SPO2 fine tune process
- added support both PD sampling and switch sampling
- ECG: support frequency to 0.05Hz to 500Hz for sine/ triangle/ square
- added support async pacing rate
Version 1.0.2.22- 2018-05-03
- fix ECG output lead setting is not synced between ECG page and PWV page
- fix incorrect PTTp time when playing the ECG and PPG waveforms
- add 'Model Name' in device information
- update terms - 'PWV' to 'PWTT', 'SPO2' to 'SpO2'
- update 'Max Peak' in PD sampling; show the peak value of the current drawing samples
- update Assistant YY0885 , 51.5.5 pass criteria
Version 1.0.2.5 - 2018-01-25
- update frequency spec in ECG page
- fix frequency accuracy issue
- update ECG simulated waveform
Version 1.0.2.4 - 2018-01-23
- fix AP issues
- added support firmware update feature
- added support assistant key function
- move RLD test from PWV page to ECG page
- update pd sampling function; load the pd sampling calibration data when connected
- set the default value of PPG 'Inverted' status to be true
- do not invert the displayed PPG waveform even the PPG 'Inverted' is checked
- added support PWV database player
- added support PPG database player
Version 1.0.1.3 - 2017-12-15
- fix AP crash issues
- ECG amplitude added by noise amplitude should be in the HW limitation range
- PPG amplitude added by noise amplitude should be in the HW limitation range
- update the deault waveform settings
Version 1.0.0.0 - 2017-09-25
- initial version