WhaleTeq offers EEG test solutions compliant with the performance testing requirements of IEC 60601-2-26:2012 and IEC 80601-2-26:2019. Multiple parameters allow user to test EEG performance with flexibilities. Meanwhile, user is able to download and play PhysioNet EEG Motor Movement/Imagery Database or CHBMIT Scalp EEG Database within a few clicks. On top of that, User can also play programmed or recorded waveform files. WhaleTeq EEG test solutions assist to verify EEG performance with efficiency and effectiveness.

 

IEC 60601-2-26:2012 
Test Clause SEEG 100 CMRR 3.0+
201.12.1.101.1 Accuracy of signal reproduction O  
201.12.1.101.2 Input dynamic range and differential offset voltage O  
201.12.1.101.3 Input noise O  
201.12.1.101.4 Frequency response O  
201.12.1.101.5 Common mode rejection   O

IEC 80601-2-26:2019 
Test Clause SEEG 100E CMRR 3.0E
201.12.1.102 Accuracy of amplitude and rate of variation O  
201.12.1.103 Input dynamic range and differential offset voltage O  
201.12.1.104 Input noise    O
201.12.1.105 Frequency response O  
201.12.1.106 Common mode rejection   O