The EEG performance tester designed for regulation testing. Standard Coverage : IEC 60601-2-26:2012, JJG 1043, JJF 1388, JJF 1390. Applicable for R&D, compliance and production line testing
Parameter | Specification | System capability |
---|---|---|
Main output voltage accuracy | ±1% for amplitudes of 0.2 mVpp or higher | ±0.3% |
Main output voltage resolution (DAC resolution) | 0.5μV | 0.5μV |
Frequency / pulse repetition rate accuracy | ±1% | ±0.1% |
Pulse duration / timing accuracy | ±1ms | ±0.2ms |
Resistor tolerance | ±1% | ±0.5% |
Capacitor tolerance | ±5% | ±5% |
Precision 1000:1 divider (100KΩ:100Ω ) | ±0.2% | ±0.05% |
Sample rate | 5kHz ± 0.1% | 5kHz ± 0.05% (50ppm) |
DC offset (fixed, noise free, from internal super capacitor) | 300mV ± 1% | 300mV ±0.1% |
Power supply | USB +5Vdc supply (no separate power supply required) 0.5A (high power mode) |
Typical load<0.25A, up to 0.45A is possible if all relays are turned on |
Environment | 15 ~ 30°C |
SEEG 100 Test System | |
---|---|
Model No. | Description |
SEEG 100 | Single Channel EEG Test System with one EEG Breakout Box and Assistant Software for IEC 60601-2-26:2012 performance tests. |
Optional Accessory | |
---|---|
Model No. | Description |
WUI100 | USB isolator for reducing the power noise from PC. Recommanded to use with SECG 4.0, MECG 2.0, HRS200, HRS100+ and SEEG 100. |
Optional Calibration Service and Warranty Extension | |
---|---|
Model No. | Description |
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 | Extend the limited warranty from (1) year to (3) years. |