The multifunction tester designed for verifying healthcare wearable with ECG and PPG technologies in the production line testing. Provide SDK (Software Development Kit) for user to develop customized or automated test software.
ECG Mode Parameter | Specification |
---|---|
Main Output Voltage Resolution |
2.5μV (DAC resolution) |
Main Output Voltage Accuracy |
±1% for amplitudes of 0.5mVpp or higher |
Frequency/Pulse Repetition Rate Accuracy | ±1% |
Pacing Pulse Amplitude Range | ±1mV ~ ±1000mV |
Pacing Pulse Amplitude Accurcy | ±2mV pulse: ±1% >2mV pulse,<-2mV: ±10% |
Pacing Pulse Width Range | 0.1ms~2ms |
Pacing Pulse Width Resolution | 0.1ms |
Pacing Pulse Width Accuracy | ±5μs |
Pacing Pulse Rise/Fall Time | 5μs |
Pacing Pulse Overshoot | <1% |
Pacing Pulse Setting Time | <1% |
Sample Rate | 1K (Normal mode);10K (Raw data mode) |
Sample Rate Accuracy | Depend on system clock accuracy |
DC Offset Range (fixed, noise free, sourced from internal super capacitor) |
-300mV, +300mV |
DC Offset Accuracy (fixed, noise free, sourced from internal super capacitor) |
±1% |
DC Offset Range (variable, may include up to 50μVpp noise) |
0~±500mV |
DC Offset Accuracy (variable, may include up to 50μVpp noise) |
±1% |
RCA Output Amplitude Range | ECG electrode output x1000, max 10V |
RCA Output Amplitude Accuracy | ±1% for amplitudes of 0.5Vpp or higher |
PPG Mode Parameter | Specification |
---|---|
BPM Range | 10~300BPM |
BPM Resolution | 1 BPM |
BPM Accuracy | ±1 BPM |
LED AC Level Reference Output Range (RCA) | 0~3 V(Full range) |
LED AC Level Reference Output Resolution | (1/3000) Full Range |
LED AC Level Reference Output Accuracy | ±6% |
Lumen Range |
50 ~ 925 Lux (full range) (5mm from diffusion film) |
Lumen Resolution | (1/3000) Full Range |
Lumen Accuracy | ±6% |
Light Wavelength Range | 525nm |
Light Wavelength Resolution | ±10nm |
S/N Ratio | >51dB (with USB isolator) |
PD Response Time (Rising) | 1μs Typical (Varies from DUT LED intensity) |
PD Response Time (Falling) | 1μs Typical (Varies from DUT LED intensity) |
PWTT Mode Parameter | Specification |
---|---|
Time Difference (PTTp, PTTf) Range |
0~1999ms (decrease with increased BPM) |
Time Difference (PTTp, PTTf) Resolution | 1ms |
Time Difference (PTTp, PTTf) Accuracy | ±1ms |
SpO2 Parameter | Specification |
---|---|
BPM Range | 10~300 BPM |
BPM Resolution | 1 BPM |
BPM Accuracy | ±1 BPM |
LED AC Level Reference Output Range (RCA) | 0~3V(Full range) |
LED AC Level Reference Output Resolution (RCA) | (1/3000 )full range |
LED AC Level Reference Output Accuracy (RCA) | ±6% |
PI (AC level/DC level) Range |
0.025% ~ 30% (AC/DC, varies with AC or DC level ) |
Light Wavelength Range | Red: 660nm; Infrared: 940nm or 880nm |
Light Wavelength Accuracy | Red:±10nm; Infrared: ±10nm |
S/N Ratio | >51DB (with USB isolator) |
PD Rising Response Time | 1μs Typical (Varies from DUT LED intensity) |
PD Falling Response Time Rising | 1μs Typical (Varies from DUT LED intensity) |
SpO2 (%) Range | 30%~100% |
SpO2 (%) Resolution | 1% |
SpO2 (%) Accuracy | Use first degree of polynomial to set R-curve, ±1~±3(%)@70~100(%)+specified accuracy of the DUT.Use second degree of polynomial to set R-curve,±1~±2(%)@70~100(%)+specified accuracy of the DUT |
[ATE100PE] ECG+PPG Automated Test Solution for Wearable Using Mediatek MT2511
Release Date:1970/01/01
ATE200PE Test System | |
---|---|
Model No. | Description |
ATE200PE-1R-525 | The test system includes PPG fixture base with 525nm Green LED |
ATE200PE-2R-940 | The test system includes PPG fixture base with 660nm RED LED and 940nm IR LED |
ATE200PE-2R-880 | The test system includes PPG fixture base with 660nm RED LED and 880nm IR LED |
Optional Calibration Service and Warranty Extension |
|
---|---|
Model No. | Description |
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 | Extend the limited warranty from (1) year to (3) years. |