It comes with various ECG patterns for AED defibrillation tests and an embedded battery voltage multimeter. The main features are data retrieval and routine creation. It can manage a complete history of every measurement the user has made in the device, including AED test results, date and time, along with user notes. With AIMS Cloud CMS for synchronization management, it is responsible for work order viewing, work order tracking, path planning, etc.
- AED Testing Scenario
- Defibrillator Testing Scenario (Optional)
Energy Measurement Specifications
Items | Specifications |
---|---|
Maximum Energy | Range: Up to 600J |
Voltage |
Range: Up to 5000Vp |
Load Resistance | Resistance: 50Ω |
Pulse Width |
Range: 1.0 ~ 50.0ms |
Maximum Average Power | 12W, equivalent to 1 defib pulse of 360J every 30 seconds |
Battery Level |
Measurement Voltage: 0.40 ~ 40.00V |
Battery Level |
Measurement Voltage: 3.60 ~ 40.00V |
General Specifications
Items | Specifications |
---|---|
Temperature |
Operating: 0°C ~ 50°C (32°F ~ 122°F) |
Humidity | 10% ~ 90% non-condensing |
Communications |
Wireless (Low Energy Radio) |
Operation Modes | Remote (APP) |
Power |
9V alkaline battery x 1 |
Mechanical |
Housing: ABS Plastic |
Compliance Specifications
Items | Specifications |
---|---|
Safety Standards | CE: IEC/EN61010-1:2010 + A1:2016; Pollution degree 2 61010-2-030:2017 |
Electromagnetic Compatibility Standards (EMC) |
CE: EN61326-1; EN301489‐1 / EN301489‐17 FCC: EMC P15B |
Other Standards |
CE: RF EN 300328; EN62311 NCC: RF LP0002 |
Specifications are subject to change without prior notice.
DFS200 Test System
Note:
(1) Customization of AED interface cable adapter is available. Please contact WhaleTeq (service@whaleteq.com).
(2) The battery measurement probe in this package can be replaced with the battery measurement probe with load (300-DFBMPLD). For the details and quotation, please contact WhaleTeq.
Optional Accessories
Model No. | Image | Description |
---|---|---|
K27-1800304 |
USB Transfer Cable: Type-A to Type-C (Male to Male), 1.8 meter |
|
G34-0600101 | ![]() |
DFS carrying bag |
300-DFBMPLD | ![]() |
Battery Measurement Probe with Load |
*Application note : Use DFS200 Battery Measurement Probe (with load) to check AED battery level
Optional DFS200 Paddle Box Set
Model No. | Image | Description |
---|---|---|
300-DFPDOCK |
|
DFS Tester Docking Base for Defibrillator |
E13-070P101 |
|
DFS Tester Defibrillation Receiving Plate |
Optional Calibration Service and Warranty Extension
Model No. | Description |
---|---|
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 |
Extend the limited warranty from (1) year to (3) years. |