Overview
- Designed for the latest EEG standard IEC 80601-2-26:2019
- Sophisticated electrical and shielding design enables the test system to output 1μV signal
- 9 channels with auto-select switches for user to test performance individually
- Built-in test circuits as defined in EEG standards with selection of 620kΩ/4.7nF for input impedance test
- Able to load and replay files in PhysioNet's EDF format, binary format and text format
- IEC 80601-2-26 helper simplifies the medical standard with test sequences, options, parameters and pass criterions
- Software Development Kit (SDK) assists user to develop customized or automated test software with less efforts
Specification
Parameters | Specifications |
---|---|
Main output voltage accuracy | ±1% for amplitudes of 50 µVpp or higher |
Main output voltage resolution (DAC resolution) | 0.5µV |
Frequency / pulse repetition rate accuracy | ±1% |
Pulse duration / timing accuracy | ±1ms |
Resistor tolerance | ±1% |
Capacitor tolerance | ±5% |
Precision 500:1 divider (50KΩ:100Ω) | ±0.2% |
Sample rate | 5kHz ± 0.1% |
DC offset (fixed, noise free, from internal supercapacitor) | 150mV ± 1% |
DC offset (variable, may include up to 50µVpp noise) | Setting ±1% or ±3mV |
Power supply | USB +5Vdc supply (no separate power supply required) 0.5A (high power mode) |
Environment | 15 ~ 30°C (by design, not tested) 30 ~ 80% RH (by design, not tested) |
Safety, EMC standards | No applicable safety standards (maximum internal voltages 12Vdc) For EMC no testing performed. CE marking based on careful selection of parts, including USB protection IC, as well as special filters to reduce noise from microprocessor (8MHz) and DC/DC converter (200kHz). |
Video
Ordering information
SEEG 100E Test System
Model No. | Description |
---|---|
SEEG 100E | Single Channel EEG Test System with one EEG Breakout Box and Assistant Software for IEC 80601-2-26:2019 performance tests. |
Optional Accessories
Model No. | Description |
---|---|
WUI100 | USB isolator for reducing the power noise from PC. Recommanded to use with SECG 4.0, MECG 2.0, HRS200, HRS100+, SEEG 100 and SEEG 100E. |
Optional Calibration Service and Warranty Extension
Model No. | Description |
---|---|
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 | Extends the limited warranty from (1) year to (3) years. |