AECG100
AECG100

AECG100

The Innovative simulator to output ECG and PPG analog signals simultaneously.
Adjustable time difference parameters to generate various Pulse Wave Transit Time (PWTT) signals.
Standard Coverage: IEC 60601-2-47, YY 0885, YY 9706.247.
Applicable for R&D, compliance and production line testing
Request a quote Inquiry list

DatasheetDatasheet
User ManualUser Manual

Overview

  • AECG100 main console unit includes ECG module with built-in test circuits with auto select switches per ECG standard required
  • PPG-1R-525 module offers optical waveforms simulating PPG heart rate signals
  • PPG-2TF-660, PPG-2R-880, and PPG-2R-940 modules offer red and infrared optical waveforms for SpO2 testing
  • Adjustable PWTT parameters allows user to check and improve the algorithm of blood pressure applications
  • Developing R curves of device under test (DUT) for SpO2 monitors' verification and validation testing
  • Test parameters in each test mode can be saved and integrated as a series of customized test sequences for improving the testing efficiency
  • Waveform player function reproduces and repeats recorded or programmed waveforms
  • AECG Assistant software pack simplifies test sequences of medical standards into adjustable parameters, selectable options, test procedures and pass criteria
  • Software Development Kit (SDK) assists users in developing customized or automated test software with less efforts

Related Information

Specification

ECG Mode

ParametersSpecifications
Main Output Voltage Resolution2.5μV (DAC resolution)
Main Output Voltage Accuracy±1% for amplitudes of 0.5mVpp or higher
Frequency / Pulse Repetition Rate Accuracy±1%
Pulse Duration / Timing Accuracy±1ms
Pacing Pulse Amplitude Setting Range-1000mV ~ +1000mV
Pacing Pulse Amplitude Accuracy±2mV pulse: ±1%
>2mV or <-2mV pulse: ±10%
Pacing Pulse Width Setting Range0.1ms ~ 2ms
Pacing Pulse Width Minimum Step Size0.1ms
Pacing Pulse Width Accuracy±5μs
Pacing Pulse Characteristics Rise/Fall Time5μs
Pacing Pulse Characteristics Overshoot<1%
Pacing Pulse Characteristics Setting Time<1%
Sample Rate50 ~ 40kHz (Normal mode/Standalone mode); 40kHz (Raw data mode)
Sample Rate AccuracyDepend on system clock accuracy
DC Offset Range 
(fixed, noise free, sourced from internal supercapacitor)
-300mV, +300mV
DC Offset Accuracy
(fixed, noise free, sourced from internal supercapacitor)
±1%
DC Offset Setting Range 
(variable, may include up to 50μVpp noise)
-500mV ~ +500mV
DC Offset Accuracy 
(variable, may include up to 50μVpp noise)
±1%
RCA Output Amplitude RangeECG electrode output x1000, max 10V
RCA Output Amplitude Accuracy±1% for amplitudes of 0.5Vpp or higher
S/N Ratio>51DB (with USB isolator)

 

PPG Mode

ParametersSpecifications
Heart Rate Setting Range10 ~ 300BPM
Heart Rate Minimum Step Size1 BPM
Heart Rate Accuracy±1 BPM 
LED DC Level Reference Output Setting Range100 ~ 3000mV
LED DC Level Reference Output Minimum Step Size1mV
LED AC Level Reference Output Setting Range0.75 ~ 30mV
LED AC Level Reference Output Minimum Step Size0.01mV
Lumen Setting Range50 ~ 925 Lux (full range) (5mm from diffusion film)
Lumen Resolution(1 / 3000) full range
Lumen Accuracy±6%
Light Wavelength Range525 / 660 / 880 / 940nm
Light Wavelength Accuracy±10nm
LED Scan Rate50 ~ 40kHz (Normal mode/Standalone mode); 10kHz (Raw data mode)
LED Scan Rate Accuracy±5μs
PD Sample Rate250kHz (single channel)
PD Sample Rate Accuracy±5μs
PD Response Time (Rising)1μs Typical (Varies from DUT LED intensity)
PD Response Time (Falling)1μs Typical (Varies from DUT LED intensity)

 

PWTT Mode

ParametersSpecifications
Time Difference (PTTp, PTTf) Setting Range0 ~ 5999ms (for the heart rate setting as 10BPM. As the heart rate value increases, the setting range of time difference will decrease accordingly.)
Time Difference (PTTp, PTTf) Minimum Step Size1ms
Time Difference (PTTp, PTTf) Accuracy±1ms

 

SpO2 Mode

ParametersSpecifications
Heart Rate Setting Range10 ~ 300 BPM
Heart Rate Minimum Step Size1 BPM
Heart Rate Accuracy±1 BPM
LED DC Level Reference Output Setting Range100 ~ 3000mV
LED DC Level Reference Output Minimum Step Size    1mV 
LED AC Level Reference Output Setting Range 0.75 ~ 30mV 
LED AC Level Reference Output Minimum Step Size 0.01mV 
PI (AC level/DC level) Setting Range 0.025% ~ 30% (AC/DC, varies with AC or DC level) 
PI (AC level/DC level) Accuracy NA 
LED Light Wavelength Range Red: 660nm; Infrared: 940nm/880nm 
LED Light Wavelength Accuracy Red:±10nm; Infrared: ±10nm 
Red Light Absolute Irradiance(100%) Intensity§ 3.55 mW / m^2 
Red Light Absolute Irradiance(100%) Accuracy ±15% 
Infrared Light Absolute Irradiance(100%) Intensity§ 6.65 mW / m^2 
Infrared Light Absolute Irradiance(100%) Accuracy ±15% 
LED Scan Rate50 ~ 40kHz (Normal mode/Standalone mode); 10kHz (Raw data mode)
LED Scan Rate Accuracy±5μs
PD Sample Rate250kHz (single channel)
PD Sample Rate Accuracy±5μs
PD Rising Response Time1μs Typical* 
PD Falling Response Time 1μs Typical* 
SpO2 (%) Setting Range1% ~ 100% 
SpO2 (%) Minimum Step Size 1% 
SpO2 (%) Accuracy ±1%+specified accuracy of the DUT (using Masimo Radical-7 Pulse CO-Oximeter or "Covidien" Nellcor Portable SpO2 Patient Monitoring System) 

§Data Collection Method: To connect the spectrometer with a 2-inch integrating sphere, and fix the integrating sphere on the DUT to measure the absolute irradiance of the LED of the DUT.

*Note: PD response time varies with the light intensity of the DUT.

Ordering information

AECG100 Main Console Unit 

Part No.DescriptionQuantity
100-AE00001Model No.: AECG100
The main console unit for ECG testing.
Supports ECG and Auto Sequence test modes.
1

 

PPG Module

Part No.DescriptionQuantity
100-AE00002Model No.: PPG-1R-525 
Model Name: Reflectance Green Light Heart Rate Module
The reflectance PPG module of 525nm green LED. 
Supports PWTT / PPG / Auto Sequence test modes.
1
100-AE00004Model No.: PPG-2R-880
Model Name: Reflectance SpO2 Module
The reflectance PPG module of 880nm IR LED and 660nm Red LED. 
Supports PWTT / SpO2 / PPG / Auto Sequence test modes.
1
100-AE00005Model No.: PPG-2R-940 
Model Name: Reflectance SpO2 Module
The reflectance PPG module of 940nm IR LED and 660nm Red LED. 
Supports PWTT / SpO2 / PPG / Auto Sequence test modes.
1
100-AE00007Model No.: PPG-2TF-660
Model Name: Transmittance SpO2 module
The transmittance PPG module.
Supports PWTT / Auto Test SpO2 / SpO2 / PPG / Auto Sequence test modes.
(two MCX (RF) male to BNC cables included)
1

 

Optional Software Add-on Pack

Part No.Description
HA0-AE01001Auto setup for IEC 60601-2-47:2012 performance tests.
HA0-AE01002Auto setup for YY 0885-2013 performance tests.
HA0-AE01003Auto setup for YY 9706.247-2021 performance tests.
HC0-AE01001The WhaleTeq database includes 30 records of PPG waveforms recorded in clinical trials to use in PPG test mode and 10 records of PPG and ECG waveforms recorded in clinical trials to use in PWTT test mode. The 25 recorded disease waveforms include symptoms such as AF (Atrial Fibrillation), APC (Atrial Premature Complexes), VPC (Ventricular Premature Contraction) and first degree AV block.
HC0-AE01002Standard waveforms in ECG, PPG, SpO2 and PWTT test modes can be saved in raw data formats for further modification

 

Optional Calibration Service and Warranty Extension

Part No.Description
YY0007Model No.: C3
Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval.
YY0008Model No.: R3
Extends the limited warranty from (1) year to (3) years.

 