Overview
- AECG100 main console unit includes ECG module with built-in test circuits with auto select switches per ECG standard required
- PPG-1R-525 module offers optical waveforms simulating PPG heart rate signals
- PPG-2TF-660, PPG-2R-880, and PPG-2R-940 module offer red and infrared optical waveforms for SpO2 testing
- Adjustable PWTT parameters allows user to check and improve the algorithm of blood pressure applications
- Developing R curves of device under test (DUT) for SpO2 monitors' verification and validation testing
- Test parameters in each test mode can be saved and integrated as a series of customized test sequences for improving the testing efficiency
- Waveform player function reproduces and repeats recorded or programmed waveforms
- AECG Assistant software pack simplifies test sequences of medical standards into adjustable parameters, selectable options, test procedures and pass criterion
- Software Development Kit (SDK) assists user to develop customized or automated test software with less efforts
Related Information
Specification
ECG Mode
Parameters | Specifications |
---|---|
Main Output Voltage Resolution | 2.5μV (DAC resolution) |
Main Output Voltage Accuracy | ±1% for amplitudes of 0.5mVpp or higher |
Frequency / Pulse Repetition Rate Accuracy | ±1% |
Pulse Duration / Timing Accuracy | ±1ms |
Pacing Pulse Amplitude Setting Range | -1000mV ~ +1000mV |
Pacing Pulse Amplitude Accuracy | ±2mV pulse: ±1% >2mV or <-2mV pulse: ±10% |
Pacing Pulse Width Setting Range | 0.1ms ~ 2ms |
Pacing Pulse Width Minimum Step Size | 0.1ms |
Pacing Pulse Width Accuracy | ±5μs |
Pacing Pulse Characteristics Rise/Fall Time | 5μs |
Pacing Pulse Characteristics Overshoot | <1% |
Pacing Pulse Characteristics Setting Time | <1% |
Sample Rate | 50 ~ 40K (Normal mode/Standalone mode); 40K (Raw data mode) |
Sample Rate Accuracy | Depend on system clock accuracy |
DC Offset Range (fixed, noise free, sourced from internal supercapacitor) | -300mV, +300mV |
DC Offset Accuracy (fixed, noise free, sourced from internal supercapacitor) | ±1% |
DC Offset Setting Range (variable, may include up to 50μVpp noise) | -500mV ~ +500mV |
DC Offset Accuracy (variable, may include up to 50μVpp noise) | ±1% |
RCA Output Amplitude Range | ECG electrode output x1000, max 10V |
RCA Output Amplitude Accuracy | ±1% for amplitudes of 0.5Vpp or higher |
S/N Ratio | >51DB (with USB isolator) |
PPG Mode
Parameters | Specifications |
---|---|
Heart Rate Setting Range | 10 ~ 300BPM |
Heart Rate Minimum Step Size | 1 BPM |
Heart Rate Accuracy | ±1 BPM |
LED DC Level Reference Output Setting Range | 100 ~ 3000mV |
LED DC Level Reference Output Minimum Step Size | 1mV |
LED AC Level Reference Output Setting Range | 0.75 ~ 30mV |
LED AC Level Reference Output Minimum Step Size | 0.01mV |
Lumen Setting Range | 50 ~ 925 Lux (full range) (5mm from diffusion film) |
Lumen Resolution | (1 / 3000) full range |
Lumen Accuracy | ±6% |
Light Wavelength Range | 525 / 660 / 880 / 940nm |
Light Wavelength Accuracy | ±10nm |
LED Scan Rate | 50 ~ 40K (Normal mode/Standalone mode); 10K (Raw data mode) |
LED Scan Rate Accuracy | ±5μs |
PD Sample Rate | 250K (single channel) |
PD Sample Rate Accuracy | ±5μs |
PD Response Time (Rising) | 1μs Typical (Varies from DUT LED intensity) |
PD Response Time (Falling) | 1μs Typical (Varies from DUT LED intensity) |
PWTT Mode
Parameters | Specifications |
---|---|
Time Difference (PTTp, PTTf) Setting Range | 0 ~ 5999ms (for the heart rate setting as 10BPM. As the heart rate value increases, the setting range of time difference will decrease accordingly.) |
Time Difference (PTTp, PTTf) Minimum Step Size | 1ms |
Time Difference (PTTp, PTTf) Accuracy | ±1ms |
SpO2 Mode
Parameters | Specifications |
---|---|
Heart Rate Setting Range | 10 ~ 300 BPM |
Heart Rate Minimum Step Size | 1 BPM |
Heart Rate Accuracy | ±1 BPM |
LED DC Level Reference Output Setting Range | 100 ~ 3000mV |
LED DC Level Reference Output Minimum Step Size | 1mV |
LED AC Level Reference Output Setting Range | 0.75 ~ 30mV |
LED AC Level Reference Output Minimum Step Size | 0.01mV |
PI (AC level/DC level) Setting Range | 0.025% ~ 30% (AC/DC, varies with AC or DC level) |
PI (AC level/DC level) Accuracy | NA |
LED Light Wavelength Range | Red: 660nm; Infrared: 940nm/880nm |
LED Light Wavelength Accuracy | Red:±10nm; Infrared: ±10nm |
Red Light Absolute Irradiance(100%) Intensity§ | 3.55 mW / m^2 |
Red Light Absolute Irradiance(100%) Accuracy | ±15% |
Infrared Light Absolute Irradiance(100%) Intensity§ | 6.65 mW / m^2 |
Infrared Light Absolute Irradiance(100%) Accuracy | ±15% |
LED Scan Rate | 50 ~ 40K (Normal mode/Standalone mode); 10K (Raw data mode) |
LED Scan Rate Accuracy | ±5μs |
PD Sample Rate | 250K (single channel) |
PD Sample Rate Accuracy | ±5μs |
PD Rising Response Time | 1μs Typical* |
PD Falling Response Time | 1μs Typical* |
SpO2 (%) Setting Range | 1% ~ 100% |
SpO2 (%) Minimum Step Size | 1% |
SpO2 (%) Accuracy | ±1%+specified accuracy of the DUT (using Masimo Radical-7 Pulse CO-Oximeter or "Covidien" Nellcor Portable SpO2 Patient Monitoring System) |
§Data Collection Method: To connect the spectrometer with a 2-inch integrating sphere, and fix the integrating sphere on the DUT to measure the absolute irradiance of the LED of the DUT. *Note: PD response time varies with the light intensity of the DUT. |
Video
Webinar
Tutorial

【Tutorial】Using Python to Conduct Quick Test via WhaleTeq SDK

【Tutorial】Load and Convert Digital File to Analog Waveform

【AECG100】SpO2 Quick Test for Masimo Radical 7

【AECG100】Reflectance SpO2 Test with Garmin Fenix 5X Plus

【AECG100】Single lead ECGs in Parallel Testing

【AECG100】ECG AF Test with Apple Watch 4

【AECG100】PWTT Test Demonstration

【AECG100】PPG Heart Rate Quick Test

【AECG100】Basic Instruction for ECG, PPG, PWTT and Auto Sequence Test Modes
Ordering information
AECG100 Main Console Unit
Part No. | Description | Quantity |
---|---|---|
100-AE00001 | Model No.: AECG100 The main console unit for ECG testing. Supports ECG and Auto Sequence test modes. | 1 |
PPG Module
Part No. | Description | Quantity |
---|---|---|
100-AE00002 | Model No.: PPG-1R-525 Model Name: Reflectance Green Light Heart Rate Module The reflectance PPG module of 525nm green LED. Supports PWTT / PPG / Auto Sequence test modes. | 1 |
100-AE00004 | Model No.: PPG-2R-880 Model Name: Reflectance SpO2 Module The reflectance PPG module of 880nm IR LED and 660nm Red LED. Supports PWTT / SpO2 / PPG / Auto Sequence test modes. | 1 |
100-AE00005 | Model No.: PPG-2R-940 Model Name: Reflectance SpO2 Module The reflectance PPG module of 940nm IR LED and 660nm Red LED. Supports PWTT / SpO2 / PPG / Auto Sequence test modes. | 1 |
100-AE00007 | Model No.: PPG-2TF-660 Model Name: Transmittance SpO2 module The transmittance PPG module. Supports PWTT / Auto Test SpO2 / SpO2 / PPG / Auto Sequence test modes. (two MCX (RF) male to BNC cables included) | 1 |
Optional Software Add-on Pack
Part No. | Description |
---|---|
HA0-AE01001 | Auto setup for IEC 60601-2-47:2012 performance tests. |
HA0-AE01002 | Auto setup for YY 0885-2013 performance tests. |
HA0-AE01003 | Auto setup for YY 9706.247-2021 performance tests. |
HC0-AE01001 | The WhaleTeq database includes 30 records of PPG waveforms recorded in clinical trials to use in PPG test mode and 10 records of PPG and ECG waveforms recorded in clinical trials to use in PWTT test mode. The 25 recorded disease waveforms include symptoms such as AF (Atrial Fibrillation), APC (Atrial Premature Complexes), VPC (Ventricular Premature Contraction) and first degree AV block. |
HC0-AE01002 | Standard waveforms in ECG, PPG, SpO2 and PWTT test modes can be saved in raw data formats for further modification |
Optional Calibration Service and Warranty Extension
Part No. | Description |
---|---|
YY0007 | Model No.: C3 Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
YY0008 | Model No.: R3 Extends the limited warranty from (1) year to (3) years. |