ESU Accessories Testing
The standard requires the manufacturer to test the thermal performance and contact impedance in regard to NE (Neutral Electrode). WhaleTeq developed the HFPA150, a neutral electrode impedance tester, to test as per the requirements of IEC 60601-2-2. Furthermore, two neutral electrode tests can be completed along with the HFPA150 and specified equipment.
IEC 60601-2-2
201.15.101 NEUTRAL ELECTRODES | HFPA150 |
---|---|
201.15.101.6 NE contact impedance | O |
Note: An oscilloscope shall be used with the HFPA150 to measure the output voltage and current.