Overview
- Elaborated design offers convenience and safety
- Dual module design supports different standard requirements
- Built-in impedance 51k/47nF for skin impedance simulation
- Simulate 1mV ECG signal for electrosurgical interference suppression (ESIS)
- Power rating >300W and resistor inductance <5μH
- Able to use BB-C1 breakout box bundled with SECG 4.0 and MECG 2.0
Specification
Items | Specifications |
---|---|
Standard Coverage | Built-in test circuits as defined in IEC 60601-2-25/-2-27/-2-49:2011 Clause 202.6.2.101 (Electrosurgeryinterference); China YY1079:2008 Clause 5.2.8.13; AAMI EC13:2002 Clause 5.2.9.14 ESIO/ESIS tests |
Test Fixture Power Rating | >300W |
Resistor Inductance | <5μH |
Frequency Range | DC to 100 MHz |
Use Non-Inductive Power Resistors | Resistance Tolerances to ±5%, TCR of ±100 ppm/C, Power Rating 200 W/pcs |
Skin Impedance Simulated Circuit | Resister: 51 kΩ <±5%; Capacitor: 47 nF <±5% C0G |
ECG Simulator in Module B | (defined in EC13, acting on LL, V1~V6) - Amplitude: 1 mV ±10% - Pulse width: 75 ms ±10% - Rate: 73 bpm ±5% |
Ordering information
HFCN Test Fixture
Model No. | Description |
---|---|
HFCN | High-frequency coupling network test fixture for electrosurgery interference test |
Optional Calibration Service and Warranty Extension
Model No. | Description |
---|---|
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 | Extends the limited warranty from (1) year to (3) years. |