Highlights
- Elaborated design offers convenience and safety
- Dual module design supports different standard requirements
- Built-in impedance 51k/47nF for skin impedance simulation
- Simulate 1mV ECG signal for electrosurgical interference suppression (ESIS)
- Power rating >300W and resistor inductance <5μH
- Able to use BB-C1 breakout box bundled with SECG 4.0 and MECG 2.0
Items | Specifications |
---|---|
Standard Coverage | Built-in test circuits as defined in IEC 60601-2-25/-2-27/-2-49:2011 Clause 202.6.2.101 (Electrosurgery interference); China YY1079:2008 Clause 5.2.8.13; AAMI EC13:2002 Clause 5.2.9.14 ESIO/ESIS tests |
Test Fixture Power Rating | >300W |
Resistor Inductance | <5μH |
Frequency Range | DC to 100 MHz |
Use Non-Inductive Power Resistors | Resistance Tolerances to ±5%, TCR of ±100 ppm/C, Power Rating 200 W/pcs |
Skin Impedance Simulated Circuit | Resister: 51 kΩ <±5%; Capacitor: 47 nF <±5% C0G |
ECG Simulator in Module B | (defined in EC13, acting on LL, V1~V6) - Amplitude: 1 mV ±10% - Pulse width: 75 ms ±10% - Rate: 73 ±3 bpm |
HFCN Test Fixture
Part No. | Description | Quantity |
---|---|---|
100-HF00101 | Model No.: HFCN The high-frequency coupling network test fixture for electrosurgery interference test and its standard package includes a test module A, a test module B, and a breakout box. If an additional breakout box is needed, it can be purchased separately. For details and quotation, please contact WhaleTeq. | 1 |
Optional Calibration Service and Warranty Extension
Part No. | Description |
---|---|
YY0007 | Model No.: C3 Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
YY0008 | Model No.: R3 Extends the limited warranty from (1) year to (3) years. |