Highlights
According to the IEC 60601-2-25:2011, diagnostic ECG devices must undergo digital tests of the CTS database to ensure accuracy in amplitude, interval, and duration measurements.
In cooperation with our partner, we offer the CTS database, which consists of 19 digital waveforms: 16 CAL files are calibration signals, and 3 ANE files are analytical signals. These are ideal for manufacturers and software developers to directly test and verify whether their algorithms meet the regulatory requirements.
Besides, users can utilize our CTS/CSE Database Compliance Analyzer (CDCA) software to compare test values calculated by DUT algorithms with reference values. By importing the test results into CDCA, it can determine whether the measurements comply with the standard.
Digital Tests of CTS Database (directly input digital signals into the ECG DUT to obtain test values)
Note: For analog tests of the CTS database required for ECG system verification, our MECG 2.0 ECG Database Player can be used to convert digital signals into analog format for playback and testing on ECG devices.
CTS Database
Types | Names |
---|---|
Calibration signal | CAL05000 |
CAL10000 | |
CAL15000 | |
CAL20000 | |
CAL20002 | |
CAL20100 | |
CAL20110 | |
CAL20160 | |
CAL20200 | |
CAL20210 | |
CAL20260 | |
CAL20500 | |
CAL20502 | |
CAL30000 | |
CAL40000 | |
CAL50000 | |
Analytical signal | ANE20000 |
ANE20001 | |
ANE20002 |
Note: For the applicable test items for each waveform, please refer to "Table GG.1 – Calibration and analytical ECGs" in IEC 60601-2-25:2011.
Part No. | Description |
---|---|
HE0-CD00001 | Model No.: CTS Database CAL and ANE files |