Origins
Wearable devices have made our lives more convenient, making it easy to measure various vital signs, such as heart rate, blood oxygen saturation and blood pressure. Many well-known manufacturers have also introduced new products and continued to develop more accurate products with more functions.
When communicating with our clients, we learned that to obtain more accurate test results, engineers very much hope to have a device that can flexibly adjust waveforms. They need a simulator to verify test challenges at each stage, such as signal processing, frequency response, system integration verification, etc. Therefore, the AECG100 + PPG-1R-525 Green Light Heart Rate Module came into being, which not only meets the flexible setting of PPG waveforms expected by engineers, but can also adjust the characteristic points of the waveform in a wide range and in detail.
Also, with the original ECG function of AECG100, it can simultaneously send out ECG and PPG signals, adjust the parameters of Pulse Wave Transit Time (PWTT), and thus verify the accuracy of the blood pressure measurement algorithm.
In addition to the complete functions and flexible testing mentioned above, we also learned from our customers that engineers often need to quickly verify products before mass production. The PPG-1R-525 module can play recorded raw data or customized waveforms for repeated verification to facilitate the development process. It also contains the function of simulating respiratory changes, which can be superimposed on the test to get closer to the real situation and make the test more comprehensive.
To conduct testing more efficiently, we also provide a free Software Development Kit (SDK), which enables them to develop automated testing software by themselves, flexibly expand testing capabilities, and reduce human errors that may result from repeated settings and operations. We use the PPG-1R-525 module to accurately test the heart rate function of wearable devices, ensure accuracy and consistency, and meet product specification requirements.
Features
Select
- Basic Waveform Output
- PPG Signal Generation
- Adjustable PWTT Pulse Wave Transit Time
- Saving Test Parameters
- Raw Data Playback
- Simple Automated Test
Highlights
Simulations that Are Closer to Time and Real Situations
- Provides the function of simulating respiratory changes, making the superimposed test closer to the real situation
Raw Data Playback with High Repeatability and Reproducibility
- Reproduces arrhythmia and other pathological waveforms
Software Development Kit for Flexible Development of Testing Program
- Users can develop automated testing software according to their needs to control the testing process, reducing the burden of repetitive operations during extensive testing
PPG Mode (1)
Parameters | Specifications | |
---|---|---|
Heart Rate | Setting Range | 10 ~ 300BPM |
Minimum Step Size | 1BPM | |
Accuracy | ±1BPM | |
LED DC Level Reference Output | Setting Range | 100 ~ 3000mV |
Minimum Step Size | 1mV | |
LED AC Level Reference Output | Setting Range | 0.75 ~ 30mV |
Minimum Step Size | 0.01mV | |
Lumen | Setting Range | 50 ~ 925 Lux (full range) (5mm from diffusion film) |
Resolution | (1 / 3000) full range | |
Accuracy | ±6% | |
LED Light Wavelength | Range | 525nm |
Accuracy | ±10nm | |
LED Scan Rate | Rate | 50 ~ 40kHz (Normal mode / Standalone mode) (2) ; 10kHz (Raw data mode) |
Accuracy | ±5μs | |
PD Sample Rate | Rate | 250kHz (single channel) |
Accuracy | ±5μs | |
PD Response Time | Rising | 1μs Typical (varies from DUT LED intensity) |
Falling | 1μs Typical (varies from DUT LED intensity) |
Note:
(1) Applicable when used with the AECG100 only.
(2) The standalone mode refers to the scenario where the user conducts tests using the three test modes stored in the AECG100 main unit without activating the AECG100 software.
PWTT Mode
Parameters | Specifications | |
---|---|---|
Time Difference (PTTp, PTTf) | Setting Range | 0 ~ 5999ms (for the heart rate setting as 10BPM. As the heart rate value increases, the setting range of time difference will decrease accordingly.) |
Minimum Step Size | 1ms | |
Accuracy | ±1ms |
Heart Rate Module
Part No. | Image | Description | Quantity |
---|---|---|---|
100-AE00002 | Model No.: PPG-1R-525 Model Name: Reflectance Green Light Heart Rate Module | 1 |
Compatible Products
Optional Calibration Service and Warranty Extension
Part No. | Description |
---|---|
YY0007 | Model No.: C3 Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
YY0008 | Model No.: R3 Extends the limited warranty from (1) year to (3) years. |