Highlights
- Support comparisons of TAF (Test Annotation File), which is generated by DUT algorithms from either digital or analog database signal input, to RAF (Reference Annotation File).
- The AI Sync function automatically synchronizes the TAF (Test Annotation File) and RAF (Reference Annotation File) to eliminate their time difference for optimizing comparison results
- Error data and error timing can be displayed for advanced algorithm improvement
- Intuitive software interface allows user to complete the comparison within a few mouse clicks
- Comparison result is identical with PhysioNet WFDB library software and can be exported to a report file
Category | Detail |
---|---|
Supported Standards | IEC 60601-2-47, ANSI / AAMI EC57, YY 0885 |
Supported Databases | MIT-BIH, AHA, NST, CU, ESC |
Supported Test Options – 17 Required Capabilities | • QRS Se, QRS +P • VEB Se, VEB +P, VEB FPR • VEB COU Se, VEB COU +P, VEB S_Run Se, VEB S_Run +P, VEB L_Run Se, VEB L_Run +P • RMS heart rate error • % beats missed, % N and S missed, % V missed, % F missed, Total SHUTDOWN time |
Supported Test Options – 25 Optional Capabilities | • SVEB Se, SVEB +P, SVEB FPR • SVEB COU Se, SVEB COU +P, SVEB S_Run Se, SVEB S_Run +P, SVEB L_Run Se, SVEB L_Run +P • VF EP Se, VF EP +P, VF DU Se, VF DU +P, VF FPR, VF Time • AF EP Se, AF EP +P, AF DU Se, AF DU +P, AF FPR, AF Time • ST EP Se, ST EP +P, ST DU Se, ST DU +P |
TAF Format | • Binary-MIT annotation (*.atr) • Text-MIT Label (*.txt) • Text-AHA Label (*.txt) • Text-AAMI Label (*.txt) • Text -AHA 2 fields (*.txt) • Text -AAMI 2 fields (*.txt) |
Supported Time Unit | • Sample Index (100Hz ~ 1500Hz) • Absolute Time |
Supported Report Export Format | TXT, CSV, Excel, Word |
Product Introduction
Part No. | Description |
---|---|
HB0-RDBA099 | Model No.: RDCA Basic Rhythm database comparison software for 17 capabilities of all arrhythmia algorithm requirements |
HB0-RDPR099 | Model No.: RDCA Pro Rhythm database comparison software for 17 capabilities of all arrhythmia algorithm requirements and optional 25 capabilities |
HB0-RDEN099 | Model No.: RDCA Enterprise Rhythm database comparison software for 17 capabilities of all arrhythmia algorithm requirements and optional 25 capabilities Supports comparison with other databases* besides AHA, MIT-BIH, NST, CU, and ESC |
◆ The 17 capabilities of all arrhythmia algorithm requirements include QRS, VEB, RMS, and Total SHUTDOWN time. Please refer to item 3 in ''Specification'' for details.
◆ The optional 25 capabilities include SVEB, VF, AF, and ST and exclude HRV. Please refer to item 4 in ''Specification'' for details.
*Note: The capabilities and formats of other databases need to align with those of the RDCA to perform the comparison. For details of capabilities, please refer to item 3 and item 4 in ''Specification''.